Technical Information

Specifications F800
Excitation voltage DC 10 V±5%, Output current; within 120 mA Remote sense type(Up to 4 350 Ω load cells can be connected in parallel)
Zero adjustment range 0 to approx. 2 mV/V
Rough adjustment: Digitally controlled by rough adjustment circuit.
Fine adjustment: Automatic adjustment by digital processing.
Span adjustment range 0.3 to 2.0 mV/V
Rough adjustment: Digitally controlled by rough adjustment circuit.
Fine adjustment: Automatic adjustment by digital processing.
Min. input sensitivity 0.3 μV/count
Accuracy Non-linearity: within 0.01% FS (Typ. 0.005% FS at room temperature)
Zero drift: within 0.1 μV/°C RTI (Typ. 0.08 μV/°C)
Gain drift: within 15 ppm/°C (Typ. 5 ppm/°C)
Noise: within 0.1 μVp-p RTI (0.1 to 10 Hz)
Analog filter Bessel type low-pass filter (-12 dB/oct.),
Selectable from 2, 4, 6, 8 Hz
A/D converter Speed: 100 times/sec.(10 mS)
Resolution: 16 bit (binary)
Min. display resolution 1/10000
Secondary calibration Calibration can be carried out without an actual load by connecting a resistor to one of the bridges of a loadcell.